*TST? query
Tests the analyzer hardware and returns the results.
Query Syntax: *TST?
Example Statements: OUTPUT 711;"*TST?"
OUTPUT 711;"*tst?"
Return Format: Integer
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: confirmed
Description:
The analyzer’s self-test performs a full calibration and then compares the calibration results to specified
limits. If the results are within specified limits, the analyzer returns 0. If the results exceed the specified
limits, the analyzer returns 1.