CALCulate[1|2|3|4]:CFIT:FSCale command/query
Specifies the frequency scaling used in the curve fit operation.
Command Syntax: CALCulate[1|2|3|4]:CFIT:FSCale <number>|<step>|<bound>
<number> ::= a real number (NRf data)
limits: 1e-6:1e6
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements: OUTPUT 711;"calculate2:cfit:fscale 604144"
OUTPUT 711;"Calc2:Cfit:Fscale 582699"
Query Syntax: CALCulate[1|2|3|4]:CFIT:FSCale?
Return Format: Real
Attribute Summary: Option: 1D3 Curve Fit/Synthesis
Synchronization Required: no
Preset State: +1.0
SCPI Compliance: instrument-specific
Description:
The analyzer scales the frequency axis (the X-axis) by f/frequency scale, where f is frequency in Hertz.
Note This command must be sent before CALC:CFIT. This command is not trace specific. It
ignores the trace specifier.
CALCulate