CALCulate[1|2|3|4]:CFIT:WEIGht:AUTO command/query
Determines the weighting function used in the curve fit operation.
Command Syntax: CALCulate[1|2|3|4]:CFIT:WEIGht:AUTO OFF|0|ON|1
Example Statements: OUTPUT 711;"calc4:cfit:weig:auto ON"
OUTPUT 711;"Calculate:Cfit:Weight:Auto ON"
Query Syntax: CALCulate[1|2|3|4]:CFIT:WEIGht:AUTO?
Return Format: Integer
Attribute Summary: Option: 1D3 Curve Fit/Synthesis
Synchronization Required: no
Preset State: ON (+1)
SCPI Compliance: instrument-specific
Description:
If CALC:CFIT:WEIGHT:AUTO is ON (the default value), the curve fit operation automatically
generates a weighting function. It stores the result in the curve fit weight register. See
CALC:CFIT:WEIG:REG command for more information about the curve fit weight register.
If CALC:CFIT:WEIGHT:AUTO OFF is sent, the curve fit operation uses the weighting function stored in
the specified register. The curve fit operation will abort if the curve fit weight register is empty or
contains invalid data.
Note This command is not trace specific. It ignores the trace specifier.
CALCulate
6-18

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals