This command loads all segments of a limit. Each segment must consist of a start value
(start_X-axis_value, start_Y-axis_value) and a stop value (stop_X-axis_value, stop_Y-axis_value).
The analyzer does not clear the previous lower limit definition when you send new segments. It only
overwrites those portions of the limit redefined by the new segments. Send CALC:LIM:LOW:CLE to
clear the previous limit.
CALCulate
6-40