*RST command
Executes a device reset.
Command Syntax: *RST
Example Statements: OUTPUT 711;"*RST"
OUTPUT 711;"*rst"
Attribute Summary: Option: not applicable
Synchronization Required: yes
Preset State: not applicable
SCPI Compliance: confirmed
Description:
This command returns the analyzer to a reset state. In addition, *RST cancels any pending *OPC
command or query.
The reset state is similiar to the preset state. The preset state of each command is listed in the Attribute
Summary. In some cases, however, a command’s reset state differs from its preset state. These
commands (and their reset states) are listed below.
CALibration:AUTO is set to 0 (OFF).
FREQuency:SPAN:LINK is set to CENTer.
FREQuency:RESolution:AUTO is ON.
CALCulate:GDAPerture:APERture is set to 0.25 PCT.
HCOPy:PAGE:DIMensions:USER:LLEFt is set to +332, +1195.
HCOPy:PAGE:DIMensions:USER:URIGht is set to +9155, +7377.
Note This command is not equivalent to a front panel preset. Send SYST:PRES if you want
to send a command that is equivalent to a front panel preset.
Common Commands
3-12