ARM:RPM:THReshold command/query
Specifies the starting RPM value.
Command Syntax: ARM:RPM:THReshold {<number>[<unit>]}|<step>|<bound>
<number> ::= a real number (NRf data)
limits: 5:491520
<unit> ::= [RPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements: OUTPUT 711;"ARM:RPM:THR 1000"
OUTPUT 711;"arm:rpm:threshold 8000"
Query Syntax: ARM:RPM:THReshold?
Return Format: Real
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: +600 RPM
SCPI Compliance: instrument-specific
Description:
This command is only valid when the Start RPM Arming qualifier is enabled with the ARM:RPM:MODE
command.
The starting value is either the lowest RPM used for devices with increasing RPM (ARM:RPM:MODE
UP) or the highest RPM for devices with decreasing RPM (ARM:RMP:MODE DOWN).