ARM:TIMer command/query
Specifies the size of the step used in time step arming.
Command Syntax: ARM:TIMer {<number>[<unit>]}|<step>|<bound>
<number> ::= a real number (NRf data)
limits: 0.0:500000
<unit> ::= [S]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements: OUTPUT 711;"arm:tim 3600"
OUTPUT 711;"ARM:TIMER 1.5"
Query Syntax: ARM:TIMer?
Return Format: Real
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: +0.5 S
SCPI Compliance: confirmed
Description:
This command is only valid with time step arming (ARM:SOUR TIM).
The default unit is seconds.