ARM[:IMMediate] command
Arms the trigger if ARM:SOUR is MAN.
Command Syntax: ARM[:IMMediate]
Example Statements: OUTPUT 711;":ARM"
OUTPUT 711;"arm:imm"
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: not applicable
SCPI Compliance: confirmed
Description:
Two conditions must be met before this command arms the trigger.
Manual arming must be selected (ARM:SOUR is MAN).
Bit 1 (Settling) or bit 6 (RDY for ARM) of the Operation Status condition register must be set to 1.
ARM:IMM is ignored at all other times.
For more information about how to use status registers, see “Programming the Status System” in the
GPIB Programmer’s Guide.
ARM
5-2