CALCulate[1|2|3|4]:ACTive command/query
Selects the active trace(s).
Command Syntax: CALCulate[1|2|3|4]:ACTive A|B|C|D|AB|CD|ABCD
Example Statements: OUTPUT 711;"CALC4:ACTIVE C"
OUTPUT 711;"calc4:active ABCD"
Query Syntax: CALCulate[1|2|3|4]:ACTive?
Return Format: A|B|C|D|AB|CD|ABCD
Attribute Summary: Option: not applicable
Synchronization Required: no
Preset State: trace A
SCPI Compliance: instrument-specific
Description:
In addition to the individual traces, you can set multiple traces to be active at any time.
The trace pairs of A/B or C/D can be active as well as all four traces—A/B/C/D. If A/B or A/B/C/D are
active, trace A is considered the “most active trace.” If traces C/D are active, trace C is considered the
“most active trace.”
The trace pairs of A/C or B/D
cannot
be active at the same time.
The trace specifier, CALCULATE[1|2|3|4], is not used.
CALCulate
6-3

Table of Contents

Other manuals for Agilent Technologies 35670A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 35670A and is the answer not in the manual?

Agilent Technologies 35670A Specifications

General IconGeneral
BrandAgilent Technologies
Model35670A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals