For polar diagrams, the data table consists of four columns. The first column contains the label, the
second column contains the independent value (frequency), the third column contains magnitude, and the
fourth column contains phase. This command returns the magnitude and phase values. Use the
“CALC:MARK:X? UNIT” command to determine the unit associated with the magnitude values. Use
the “CALC:MARK:Y? UNIT” command to determine the unit associated with the phase values.
For orbit diagrams, the data table consists of four columns. The first column contains the label, the second
column contains the independent value (time), the third column contains the X-axis value, and the fourth
column contains the Y-axis value. This command returns the X-axis and Y-axis values. Use the
“CALC:MARK:X? UNIT” command to determine the unit associated with the X-axis values. Use the
“CALC:MARK:Y? UNIT” command to determine the unit associated with the Y-axis values.