Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 111
Maintenance 2
To change HPLC-Chip capillaries
• port 1 - not used
• port 2 - 15 µm (orange) to nanopump
• port 3 - 75 µm (blue) to infusion pump
• port 4 - not used
• port 5 - 100 µm (black) to waste
• port 6 - 25 µm (yellow) to port 6 of the µ-WPS
For backflush mode
• port 1 - not used
• port 2 - 15 µm (orange) to nanopump
• port 3 - 75 µm (blue) to infusion pump
• port 4 - not used
• port 5 - 25 µm (yellow) to port 6 of the µ-WPS
• port 6 - 100 µm (black) to waste
8 Route the capillaries exactly as shown in Figure 53.
.
9 Reinstall the capillary cover. Do not overtighten the cover screws!
Figure 53 Routing capillaries