EasyManua.ls Logo

Agilent Technologies 6500 Series - Page 111

Agilent Technologies 6500 Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 111
Maintenance 2
To change HPLC-Chip capillaries
port 1 - not used
port 2 - 15 µm (orange) to nanopump
port 3 - 75 µm (blue) to infusion pump
port 4 - not used
port 5 - 100 µm (black) to waste
port 6 - 25 µm (yellow) to port 6 of the µ-WPS
For backflush mode
port 1 - not used
port 2 - 15 µm (orange) to nanopump
port 3 - 75 µm (blue) to infusion pump
port 4 - not used
port 5 - 25 µm (yellow) to port 6 of the µ-WPS
port 6 - 100 µm (black) to waste
8 Route the capillaries exactly as shown in Figure 53.
.
9 Reinstall the capillary cover. Do not overtighten the cover screws!
Figure 53 Routing capillaries

Table of Contents

Related product manuals