Agilent 6500 Series Q-TOF LC/MS Maintenance Guide 111
Maintenance 2
To change HPLC-Chip capillaries
port 1 - not used
port 2 - 15 µm (orange) to nanopump
port 3 - 75 µm (blue) to infusion pump
port 4 - not used
port 5 - 100 µm (black) to waste
port 6 - 25 µm (yellow) to port 6 of the µ-WPS
For backflush mode
port 1 - not used
port 2 - 15 µm (orange) to nanopump
port 3 - 75 µm (blue) to infusion pump
port 4 - not used
port 5 - 25 µm (yellow) to port 6 of the µ-WPS
port 6 - 100 µm (black) to waste
8 Route the capillaries exactly as shown in Figure 53.
.
9 Reinstall the capillary cover. Do not overtighten the cover screws!
Figure 53 Routing capillaries