Power Measurement How to Measure Power
82 Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
If the measured power is more than 40 dBm less than the range setting, it is
impossible for power to be displayed. The power value, -1.--- dBm, as shown
in Figure 51, is shown. This means that the measured power is greater than the
resolution at this [Range] value. You must increase the [Range] value.
See Table 1 on page 83 for more details.