14 Chapter 1
Troubleshooting
Before You Start
Replacement Assemblies
The instrument assemblies are not repairable to the component level.
Refer to Chapter 3 to determine how to disassemble and assemble the
instrument. The following assemblies must be replaced as an assembly.
A1 keyboard (does not include keypads)
A2 flat panel display
A3 front panel interface assembly (includes inverter boards)
A4 disk drive assembly
A6 power supply assembly
A7 baseband I/Q assembly (Option B7C)
A10 digital IF assembly
A12 analog IF assembly
A14 fan control board
A15 daughter board
A17 RF assembly
A18 reference assembly
A19 synthesizer assembly
A21 motherboard
A22 CPU (processor) assembly (does not include A22A1, -A2, or -A3)
A22A1 DRAM card
A22A2 Flash memory board
A22A3 GP-IB capacitor board
A25 SCSI board
A27 rear panel connector board
A3A1, A3A2 inverter board
After an Instrument Repair
If an instrument assembly has been repaired or replaced, perform the
related adjustments and performance verification tests, using the
N7812A E4406A Calibration Application Software. The software can be
downloaded via the web at
http://www.agilent.com/find/calibrationsoftware. Licenses to use the
software for each instrument can be purchased through an Agilent
Technologies Sales and Service office. Table 1-2 on page 18 details the
computer and test equipment needed to use the software.