Chapter 4 181
Assembly Replacement
A6 Power Supply
Replacement
1. Place the power supply assembly into position in the deck. Push
down to mate the assembly with the motherboard connector.
2. Refer to Figure 4-8. Using the T-10 driver, replace the 5 screws.
Torque to 9 inch pounds.
3. Replace the instrument top brace. Refer to the “Top Brace”
replacement procedure.