Chapter 1 47
Troubleshooting
Isolating an RF, Analog IF, Digital IF, Reference, Synthesizer, or CPU Problem
Verifying the A10 Digital IF Assembly
This procedure assumes the A17 RF assembly and the A12 analog IF
assembly are functioning correctly. To verify proper RF and analog IF
functionality, refer to the “RF Assembly Quick Check” on page 40 and
the “Analog IF Assembly Quick Check” on page 43. The A10 digital IF
will digitize the 7.5 MHz 3rd and final IF in the analog to digital
converter. There is a 21.4 MHz “tapped” autoranging signal from the
A12 analog IF assembly that pre-adjusts an amplifier on the A10 digital
IF before the 7.5 MHz down-converted IF signal reaches the ADC. This
pre-adjusted amplifier gain is electronically adjusted so the power level
at the ADC input is optimized for maximum efficiency.
Digital IF Assembly Quick Check
The A10 digital IF assembly requires a 30 MHz clock from the A12
analog IF assembly or the instrument will not boot.
1. Inject a CW signal at 50 MHz at 25 dBm into the RF input of the
instrument.
2. Press
Preset on the instrument and tune to 50 MHz by pressing
Input, Input Port, RF, Frequency 50 MHz.
3. Press
Input and verify the input attenuation is set to 0 dB by viewing
the 5th softkey from the top. If it is not at 0 dB, press
Input Atten and
enter 0 dB.
4. Connect a calibrated spectrum analyzer to A10TP5.
5. The 7.5 MHz 3rd IF power level at A10TP5 should measure
22.5 dBm on the calibrated spectrum analyzer if the circuitry up to
the ADC is operating correctly.
Digital IF Assembly Detailed Troubleshooting
If the quick check power level is not correct, other items can be checked
to verify the A10 digital IF assembly.
The remaining detailed checks require the digital IF assembly to be
placed on an extender board (p/n E4406-60021). Turn the instrument
off before removing a PC board assembly from the instrument.
If the levels in the table below Figure 1-10 are incorrect, or the clock
measurements (using an oscilloscope) or dc levels in Figure 1-11 (if the
part number of the digital IF assembly in your instrument is
E4406-60006 see Figure 1-12) are incorrect, suspect that the A10 digital
IF assembly is defective.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals