52 Chapter 1
Troubleshooting
Isolating an RF, Analog IF, Digital IF, Reference, Synthesizer, or CPU Problem
Verifying the A19 Synthesizer Assembly
The main function of the A19 synthesizer assembly is to provide the 1st
L.O. to the A17 RF assembly. This is accomplished by using a variable
VCO with a fundamental frequency range of 500 MHz to 1000 MHz.
This signal will be divided, bandpassed, and doubled 2 or 4 times to
achieve the correct 1st L.O. frequency. The 1st L.O. frequency depends
on the setting of the instrument’s center frequency. In most cases the
L.O. will be 321.4 MHz higher than where the center frequency of the
instrument is tuned.
For example, if the center frequency of the instrument is set to
1000 MHz, the 1st L.O. from the A19 synthesizer assembly will be
1000 MHz + 321.4 MHz or 1321.4 MHz.
Since the maximum 1st L.O. frequency is 4000 MHz, the 1st L.O. will
shift 321.4 MHz lower in frequency when the instrument is set to a
center frequency of 3678.6 MHz.
For example, if the center frequency of the instrument is set to
3678.5 MHz, the 1st L.O. is 3678.5 MHz + 321.4 MHz = 3999.9 MHz
(almost max). As soon as the instrument is set to a center frequency of
3678.6 MHz and greater the 1st L.O. will shift to
3678.6 MHz − 321.4 MHz = 3357.2 MHz. The best location to measure
the 1st L.O. frequency and power level is when the A17 RF assembly is
placed on an extender board (p/n E4406-60042).
The A17 RF Assembly Detailed Troubleshooting provided a process for
measuring the 1st L.O. power when the instrument was tuned to
50 MHz. Power holes or defective circuitry on the A19 synthesizer
assembly can cause low gain at other frequencies. For example, if the
instrument has low gain at 1234.5 MHz, measuring the 1st L.O. power
would be a good troubleshooting step to narrow the failure to the A19
assembly.
Synthesizer Assembly Quick Check
Refer to Figure 1-8 (test point L) on page 42 to measure the 1st L.O.
Synthesizer Assembly Detailed Troubleshooting
Turn the instrument on with the A19 synthesizer on an extender board
(p/n E4406-60042) and verify the following levels. If the levels are
incorrect, suspect the A19 synthesizer assembly as being defective.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals