EasyManua.ls Logo

Agilent Technologies E4406A VSA Series - Replacement

Agilent Technologies E4406A VSA Series
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
184 Chapter 4
Assembly Replacement
A7 Baseband I/Q Assembly (Option B7C)
Figure 4-11 A7 Assembly with Baseband I/Q Rear Ejector
Replacement
Step 1. Insert the A7 assembly into the instrument
1. Slide the A7 assembly into the front and rear guides of slot 0 ( Note:
if the A7 assembly was retrofitted into the instrument, the A7
assembly may be located in slot 1; in this case the A7 assembly will
have a standard rear ejector, not the large rear ejector shown as item
1 in Figure 4-11).
2. Carefully insert the baseband I/Q assembly ejectors into from the
motherboard.
3. Hook the front and rear ejectors under the tabs that are located on
the rear frame and mid web of the instrument.
Step 2 Attach the Baseband I/Q Cable Assembly
1. Connect the I, I
, Q, and Q cables to their respective connectors on the
top-front of the A7 assembly, if they are not already connected.
2. Attach the four, bundled I/Q wires to the cable clip (1400-3014) that
is located on the rear of the front panel assembly, as shown in item 5
of Figure 4-9.
3. Insert the I, I
, Q and Q cables into their respective openings in the
rear of the front panel assembly. These openings are labelled.
4. Use the 9/16 deep socket nut driver to secure the four 15/32-32 hex
nuts (2950-0035) to the I, I
, Q and Q connectors on the front panel.
See item 1 in Figure 4-10.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Related product manuals