174 Chapter 4
Assembly Replacement
Top Brace
Replacement
1. Carefully position the top brace on the deck. Reference the two
alignment pins and the two alignment holes. Make sure that no
coaxial cables will get pinched underneath the brace.
2. Using the T-10 driver, replace the 7 top screws first, and then the 10
side screws after the top screws are tightened. Torque to 9 inch
pounds.
3. Replace the outer case. Refer to the “Instrument Outer Case”
replacement procedure.