EasyManuals Logo
Home>Agilent Technologies>Test Equipment>E4406A VSA Series

Agilent Technologies E4406A VSA Series Service Guide

Agilent Technologies E4406A VSA Series
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #176 background imageLoading...
Page #176 background image
176 Chapter 4
Assembly Replacement
Front Frame
Figure 4-4 Front Frame Assembly Removal
Replacement
1. Place the front frame assembly in front of the deck.
2. Connect the ribbon cable (3) to the A3 front panel interface board.
3. Feed the coaxial cable BNC connector through the External Trigger
Input
hole in the front frame. Secure with the nut removed earlier,
using a 9/16’” socket. Torque to 21 inch pounds.
4. Clip the coaxial cable into the two cable clamps positioned on the
front frame shield.
5. Position the front frame on the deck using the alignment bosses on
the deck (5). Remember to tuck the ribbon cable under the fans
when pushing the frame onto the deck. This will insure proper
airflow to cool the instrument. Using the T-10 driver, replace the 6
screws (1) that secure the front frame to the deck. Torque to 9 inch
pounds.
6. Replace the instrument outer case. Refer to the “Instrument Outer
Case” replacement procedure.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals