214 Chapter 4
Assembly Replacement
RF Input Connector
RF Input Connector
Removal
1. Remove the instrument outer case. Refer to the “Instrument Outer
Case” removal procedure.
2. Remove the front frame. Refer to the “Front Frame” removal
procedure.
3. Refer to Figure 4-29. Using a 5/16 inch wrench, disconnect the
semi-rigid cable (1) from the attenuator/connector.
4. Using the T-10 driver, remove the 2 screws (2) that secure the input
connector bracket to the deck.
5. Remove the attenuator/connector from the instrument.
6. Separate the attenuator (3) from the input connector (5) using a
5/16 inch wrench.
7. Remove the connector from the bracket (4) using a 9/16 inch
wrench.
Figure 4-29 RF Input Connector