Chapter 4 225
Assembly Replacement
Front Frame Subassemblies
Replacement
1. Place the front panel interface board into position in the front frame
assembly. Mate the connector with the connector in the keyboard.
2. Using the T-10 driver, replace the 7 screws that secure the board to
the front frame. Torque to 9 inch pounds.
3. Plug the display flat flex cable into the front panel interface board
connector. Push down on both sides of the locking mechanism.
4. Reconnect both of the 2-wire backlight cables.
5. Press the RPG knob onto the control shaft.