Chapter 1 33
Troubleshooting
Auto-Align Tests
NOTE Additional auto-alignment procedures are performed for instruments
equipped with base band I/Q inputs (Option B7C).
Nominal System Gain With the analog IF preFilter OFF and IF BW set to 5 kHz, alignment is done using the
50 MHz ref cal signal to measure the residual amplitude error (offset) at the center
frequency in default (reference) settings (eg. 10 dB atten). This error is used to compensate
acquisition data whenever there is an “absence” of a valid current system gain alignment; in
other words, this becomes the “default” gain correction to fall back on.
321 MHz Amplitude The 321 MHz cal osc signal from the A17 RF assembly is used for many internal
alignments. This alignment merely calibrates the amplitude control dac for this signal, for
use by the other alignments when necessary to set fairly accurate amplitude levels.
LC Prefilter Wide This alignment is run once for WIDE (BW>1.2 MHz) and once for NARROW. The 321 MHz
Comb cal osc is used to align BW and CF for the analog IF LC path. Coefficients are
generated for a curve that maps BW-Dac to requested BW. Another curve is generated
which represents Gain vs BW. Both these curves are then used during Prefilter and IF gain
setup during a measurement.
LC Prefilter Narrow See description for LC Prefilter Wide.
XTAL Prefilter Wide This alignment is run once for WIDE (BW>20 kHz) and once for NARROW. The 321 MHz
Comb cal osc is used to align BW, SYM, & CF for the analog IF XTAL path. Coefficients are
generated for a curve that maps BW-Dac to requested BW. Another curve is generated
which represents Gain vs BW. Both these curves are then used during Prefilter and IF gain
setup during a measurement.
XTAL Prefilter Narrow See description for XTAL Prefilter Wide.
ADC Autorange The ADC autoranging subsystem has several hardware components that need to be setup.
This alignment takes care of 3 dacs: 2 dacs on the A10 digital IF assembly that set the
positive and negative thresholds for range switches; and 1 dac on the A12 analog IF
assembly that sets the GAIN of the autoranging detector signal. It also generates 4 gain
values to be used for 4 possible measurement setups. The 4 setups cover the on/off positions
of dither and analog IF post-filter. This is necessary since these positions produce different
relative signal levels between the main path and the autorange path.
Attenuator Steps Align the relative error of each attenuator step at 50 MHz. The 10 dB attenuator setting is
considered the reference, with 0 dB error by default. A CW signal is injected at 50 MHz and
each attenuator step is measured to yield an error relative to the step’s assumed value (for
example: an attenuator setting of 21 is expected to produce 21 dB of attenuation... if it
produces 21.12, the error is 0.12 dB).
Trigger Delays This alignment currently is used ONLY for delay compensation within the CDMA
personality for measurements which synchronize to an “even second clock”. This alignment
measures the delay of the IF path for the one particular setup used for this CDMA
application.
Measure Comb Teeth The amplitude of the internal 50 MHz reference is compared to the amplitude of the 20
comb teeth (used during IF flatness alignment) by tuning the LO. If the resulting
comparison is greater than a particular amount, the alignment fails. Mainly used for wide
span comb settings that require measurements in order to achieve flatness down to 0.1 or
0.2 dB in the 10 MHz span.
Table 1-3 Sequence for Auto-Align Procedures
Procedure Name Procedure Description

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals