142  FieldFox User’s Guide  
o  2-Port Transmission 2-port transmission measurement. Learn how to make 
this measurement on page 146.  
o  B/A and A/B 2-port ratioed receiver measurements using external sources. 
Learn more on page 147.  
Frequency Selection 
VVM measurements are made at a single CW frequency. 
How to set Frequency 
NOTE  Frequency can be set to 1 Hz resolution. 
  Press Freq/Dist  
  Then Frequency 
  Enter a value using the numeric keypad, the ▲|▼ arrows, or the rotary knob. 
  After using the  ▲|▼ arrows or the rotary knob, press Enter. The increment 
setting of the arrows is based on the current span and can NOT be changed in 
VVM Mode. 
  After using the keypad, select a multiplier key. Learn about multiplier 
abbreviations on page 19.   
Display Resolution 
You can display either one digit or two digits after the decimal point for both 
magnitude and phase readings. 
This setting survives a Preset and Mode Preset. 
How to set Resolution 
  Press Scale/Amptd  
  Then Resolution   
  Then choose:  
o  1  0.0 (default)  
o  2  0.00  
IF Bandwidth 
Reducing the IF receiver bandwidth reduces the effect of random noise on a 
measurement. However, narrower IF bandwidths cause longer measurement 
times.  
For highest accuracy, Zero AFTER changing the IF Bandwidth. Learn about 
Zeroing on page 144. 
How to set IF BW  
  Press BW 2  
  Then IF BW  
  Then choose from the following: