Time Domain - Option 010 61
For all filter shapes T
2
is equal to T
3
, and the filter is the same on both sides of
the center time.
Minimum gate span is twice the cutoff time. Each gate shape has a minimum
recommended gate span for proper operation. This is a consequence of the finite
cutoff rate of the gate. If you specify a gate span that is smaller than the
minimum span, the response will show the following effects:
distorted gate shape that has no passband
distorted shape
incorrect indications of start and stop times
may have increased sidelobe levels.