84 FieldFox User’s Guide
Trace 1 - Corrected trace with antenna factor. (Antenna = ON, Apply Corr = ON)
Trace 2 - (View) Uncorrected trace (Apply Corr = OFF)
Red Trace - Current correction factor. See below.
Learn how to set Y-Axis Units on page 80.
Use a Band Power marker to measure total power over a range of frequencies.
Learn how on page 158.
How to select correction for Field Strength measurements
The Antenna and Cable correction data survives a Mode Preset and Preset.
All Correction ON/OFF states survive a Mode Preset, but NOT a Preset.
Press Scale/Amptd.
Then More
Then Corrections
Then choose from the following:
o Apply Corrections ON OFF Turn ON and OFF correction for all settings
o Antenna ON OFF Turn ON and OFF Antenna correction.
o Edit/Save/Recall Antennas Recall or edit an Antenna Correction table. See
below.
o Cable ON OFF Turn ON and OFF Cable correction.
o Edit/Save/Recall Cables Recall or edit a Cable Correction table. See below.
o View Corrections ON OFF Turn ON and OFF Red Trace - Antenna/Cable
corrections.
Using the Antenna/Cable Editor
The Antenna Editor and the Cable Editor menus are very similar. Both tables
include header information, and a Frequency/Value table.
FieldFox Antenna Editor