8. Operation Screens
8-114
ERROR RESET: Closing this terminal cancels error display and error output.
MEAS BEFORE WELD: At the point when this terminal is closed, the
measured value of displacement, force or external input before welding start
is established
DIST 0 RESET: At the point when this terminal is closed, a zero reset of
displacement measurement is performed.
FORCE 0 RESET: At the point when this terminal is closed, a zero reset of
force measurement is performed.
No setting
(2) INPUT 7 to 8
Select from the followings. Available only for the force/displacement-equipped
specification.
DIST TRIGGER: Used to start displacement measurement from external
device.
FORCE TRIGGER: Used to start force measurement from external device.
EXTERNAL TRIGGER: Used to start external voltage/current input
measurement from external device.
No setting
(Caution) Do not make INPUT 1 to INPUT 8 the same setting. If INPUT 7 and
INPUT 8 are the same, INPUT 7 works.
(3) Function keys
MENU: Touching this displays the MENU screen.
NEXT: Touching this displays the EXT INPUT/OUTPUT (1) screen.
CHECK: Touching this displays the INPUT/OUTPUT CHECK screen.