6.2.1.2 - Integrate on Chip (IOC) Enabled
6.2.1.2.1 - Default: One Pulse per Trigger
Description
This is the standard mode of operation in external trigger. Every external trigger within an exposure will generate only
one gate pulse. A gate delay is applied that is the same per every trigger within an exposure.
Setup
IOC enabled, output A/B/C disabled.
Waveforms
Operation
External Trigger
Fire
Gate Monitor
ARM
One Pulse per Trgger