6.2.2.2.3 - Gate Step Enabled
Description
The external trigger generates one gate pulse per exposure. A gate delay is applied that is the same for each exposure
within an accumulated image. The gate delay is incremented for every accumulated image that is returned.
Setup
Gate step enabled, output A/B/C disabled.
Waveforms
Operation
External Trigger
Fire
Gate Monitor
ARM