New iStar ICCD
Operation
6.2.2.2 - Number of Accumulations > 1
6.2.2.2.1 - Standard Operation (IOC Disabled and Gate Step Disabled)
Description
The external trigger generates one gate pulse per exposure. A gate delay is applied that is the same for each exposure
within an accumulated image. The gate delay is also the same for every accumulated image that is returned.
Setup
IOC disabled, gate step disabled, output A/B/C disabled.
WAVEFORMS
External Trigger
Fire
Gate Monitor
ARM