Assembly Removal and Replacement, MS202xC 6-15 Replacing Rotary Knob – 3-61360-2
MS20xxC MM PN: 10580-00307 Rev. D 6-23
6-15 Replacing Rotary Knob – 3-61360-2
This procedure covers removing and replacing the Rotary Knob. The case does not need to be opened for this
procedure. Note that this part does not include the Rotary Encoder. To replace the Rotary Encoder, refer to
Section 6-20 “Replacing Rotary Encoder – 3-410-101” on page 6-41
Part Numbers
3-61360-2 – Rotary Knob
For MS2026C, MS2027C or MS2028C instruments
Does not include the 3-410-101 – Rotary Encoder
Procedure
1. Use a medium flat blade screwdriver and a small protective piece of rubber (as a fulcrum) to remove the
Rotary Knob.
2. Position the rubber piece next to the Knob, and slide in the screwdriver between the Knob and the Case.
3. Gentle lever the Knob up and off of its mounting shaft.
4. Position the replacement Knob over the shaft and press into place. There is no required orientation of the
Knob.

Table of Contents

Other manuals for Anritsu VNA Master MS2026C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu VNA Master MS2026C and is the answer not in the manual?

Anritsu VNA Master MS2026C Specifications

General IconGeneral
BrandAnritsu
ModelVNA Master MS2026C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals