Assembly Removal and Replacement, MS202xC 6-17 Replacing Power Monitor PCB Assembly, Option 5 –
MS20xxC MM PN: 10580-00307 Rev. D 6-29
Replacing Case Rear Assembly and Battery
1. Replace the Case Bottom Assembly, making sure that the top connector panel is correctly centered in its
mounting groove in the case back. Insert and tighten the four case mounting screws.
Torque these screws to 7.5 lbf·in (0.85 N·m).
Installing Battery:
Refer to Section 5-4 “Battery Pack Removal and Replacement” on page 5-3.
2. Install the Battery.
3. Install the Battery Door.
Restarting the Instrument:
4. Restart the instrument.
5. Perform a function check to verify all repairs and part replacements.