6-16 Opening the Instrument Case Assembly Removal and Replacement, MS202xC
6-26 PN: 10580-00307 Rev. D MS20xxC MM
Closing the Case
1. With the Case Front Assembly face down on a clean, stable work surface that will not scratch the front
panel, place the Case Rear Assembly over the front and align all parts.
2. Carefully close the case, guiding the top connector panel into case grooves.
3. Use a Phillips screwdriver to replace the four (4) Pan Head Phillips-head screws securing the two Case
halves together. Torque these screws to 7.5lbf·in (0.85N·m).
4. Replace the battery and battery door. Refer to Section 5-4 “Battery Pack Removal and Replacement”
on page 5-3.
Installing Battery:
Refer to Section 5-4 “Battery Pack Removal and Replacement” on page 5-3.
5. Install the Battery.
6. Install the Battery Door.
Restarting the Instrument:
7. Restart the instrument.
8. Perform a function check to verify all repairs and part replacements.