2-4 S-Parameter Measurements Verification Vector Network Analyzer Performance Verification
2-8 PN: 10580-00307 Rev. D MS20xxC MM
If Verification Fails
If the verification fails, then check the quality, cleanliness, and installation methods for the calibration and
verification components. Specifically, check:
The VNA test port connectors
The calibration tee
The impedance transfer standards
The test port cables, for damage and cleanliness
The test port cables, for proper connection and torquing
The test port cables, for phase stability
These are the most common causes for verification failures.