7-24 Replacing Clear Plastic LCD Protector – 3-61368 Assembly Removal and Replacement, MS203xC
7-66 PN: 10580-00307 Rev. D MS20xxC MM
7-24 Replacing Clear Plastic LCD Protector – 3-61368
This procedure is written with the assumption that the Main PCB Assembly has been removed from the
instrument as described in Section 7-21 and that the Clear Plastic LCD Protector must be replaced. The
replacement procedure is the same for MS2036C, MS2037C and MS2038C instruments.
Part Numbers
• 3-61368 – Clear Plastic LCD Protector for MS2036C, MS2037C and MS2038C instruments
Procedure
1. Open the instrument case and remove the Main PCB Assembly.
Refer to Section 7-21 “Removing MS203xC Main PCB Assembly” on page 7-54
2. Remove the Clear Plastic LCD Protector by pushing it from the front.
3. The protector has an inset edge which goes into the hole and faces out. The protector “lip” is set towards
the inside of the case.
4. The outside of the Protector has a protective “skin” that should be left in place until the instrument is
reassembled.
5. Clean the inside of the Protector by using compressed air and LCD compatible wipes.
6. Avoiding fingerprints, install the protector into the Case Front opening. The protector “lip” goes towards
the inside of the case.
7. If no further unit replacements are required, then install the Main PCB and reassemble the instrument
by following the procedure in Section 7-25 “Installing Main PCB and Reassembling Instrument”
on page 7-67.
Installing Battery:
Refer to Section 5-4 “Battery Pack Removal and Replacement” on page 5-3.
8. Install the Battery.
9. Install the Battery Door.
Restarting the Instrument:
10. Restart the instrument.
11. Perform a function check to verify all repairs and part replacements.