3-5 Single Side Band (SSB) Phase Noise Spectrum Analyzer Performance Verification
3-10 PN: 10580-00307 Rev. D MS20xxC MM
51. Press the Delta On/Off soft key to turn Off Delta Marker.
52. Press the Span hard key.
53. Enter 22 and then press the MHz soft key to set Span to 22 MHz.
54. Wait until the Trace Count displays 3/3.
55. Press the Marker hard key and then the Peak Search soft key.
56. Press the Delta On/Off soft key to turn On Delta Marker.
57. Enter 10 and press the MHz soft key to set Delta frequency to 10 MHz.
58. Record the Marker readout value into the Measured Value column of Table A-14 for MS2036C, of
Table A-35 for MS2037C or of Table A-59 for MS2038C in Appendix A.
59. Subtract 30 dB from the Measured Value and record the result to the Calculated Value column of
Table A-14 for MS2036C, of Table A-35 for MS2037C or of Table A-59 for MS2038C in Appendix A.
60. Verify that the calculated result is less than the specification for 10 MHz Offset.