C.6 Option 06: Parallel BCD Output (1 millisecond Resolution) 121
Analog Signal Outputs
The two analog signal outputs can be individually enabled and configured, using the following
steps:
To enable analog signal output number 1, set jumper JMP4 to position ”B”. This setting also
eliminates the possibility of having parallel data bit 44, digital signal output number 3.
To select the type of signal for analog output 1, remove the jumper from jumper strip JMP9 (which
is no longer active, since it corresponds to digital signal output number 3), and place it on jumper
strip JMP8 in either position 1 or 2 (refer to Figure 2).
To enable analog signal output number 2, set jumper JMP5 to position ”B”. This setting also
eliminates the possibility of having parallel data bit 45, digital signal output number 4.
To select the type of signal for analog output 2, remove the jumper from jumper strip JMP10
(which is no longer active, since it corresponds to digital signal output number 4), and place it on
jumper strip JMP7 in either position 1 or 2 (refer to Figure 2).
1 Pulse-per-Second (1 PPS) Output, 1,000 Pulse-per-Second (1 kPPS) Output
The 1 PPS and 1 kPPS outputs are enabled whenever Option 06 is placed in the BCD output mode
(jumper JMP1 set to position ”B”).

Table of Contents

Other manuals for Arbiter Systems 1084A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arbiter Systems 1084A and is the answer not in the manual?

Arbiter Systems 1084A Specifications

General IconGeneral
BrandArbiter Systems
Model1084A
CategoryClock
LanguageEnglish