Service Manual i2000SR interface module
625798100.APS.5.doc Page 181 of 212
STEPS REFERENCES
18 At the i2000SR feet, use the nut to
lock the feet.
VERIFICATION
STEPS REFERENCES
1 Perform a control run on any routine
and STAT assay.
2 Verify that probe enters the sample
tube in a manner that ensures that it
does not come in contact with the
side of the tube.
3 If no error messages occur and the
relationship between the probe is
centered on the tube, the procedure
has successfully completed.
4 Reinstall the removed covers.
5 Return APS operation to users.