i2000SR interface module Service Manual
Page 180 of 212 625798100.APS.5.doc
STEPS REFERENCES
14 Place the Probe Calibration High Tool
(P/N 204022), into the IM track at the
appropriate sampling location. Ensure
that the tool is pressed firmly against
the gate.
If not, the alignment will be
compromised.
15 Follow the i2000SR screen menu
options to select the Probe Move
option.
In the probe move option both probes
are moved over the calibrated probe
location.
16 Follow the i2000SR screen menu
options to lower the probe over the
tool.
17 Verify that both the i2000SR Sample
and STAT probes are aligned
perpendicular to the tube (center of
the tool).
If the rear feet are too low, the probe
will touch the back of the tool/tube
[2]. Adjust the feet to raise the back
of the analyzer.
If the rear feet are too high, the probe
will touch the front of the tube/tool
[3]. Adjust the feet to lower the back
of the analyzer.
NOTE:
This tool is only used by an FSR.
Please ensure that this tool is stored
on site in a secure location.