i2000SR interface module Service Manual
Page 188 of 212 625798100.APS.5.doc
WARNING: Potential Biohazard. This is an activity or area
where you may be exposed to potentially infectious
material. Refer to āSection 4 - Hazards and Warnings
ā of
the Accelerator Service Manual
Reduce exposure by using proper personal protection
equipment and cleaning surfaces with 0.1% sodium
hypochlorite (bleach).
CAUTION: Moving Parts
The safety cover protection is not guaranteed if the
security devices are removed or bypassed. Use caution to
prevent injury caused by moving parts.
WARNING: Electrical Shock Hazard. Possibility of electrical
shock exists if procedural or engineering controls are not
observed.
WARNING: Probe Sharps Hazard
Follow safe sharp practices. Do not place your hand near
the probe while it is moving. Put the instrument through a
decontamination cycle and wipe the probe with a
disinfectant prior to touching or working near the probe.
WARNING: Pinch hazard
CAUTION: Class 2 Laser Radiation when open. Do not stare
into beam. Avoid eye exposure to laser light.
STEPS REFERENCES
1 Open the front panel of the i2000SR
IM electrical box and locate the
MASTER, SLAVE 0, and SLAVE 1 CAN-
BUS board 7-segments red displays.