i2000SR interface module Service Manual
Page 48 of 212 625798100.APS.5.doc
Figure 23 – i2000SR IM first alignment
To assemble the i2000SR IM, use the metal screws fixed to
the interface module. Once the analyzer and spur are positioned
in a conclusive way, all metallic screws must be replaced with
plastic screws. In the event of a seismic event (earthquake), these
screws allow the IM to breakaway from the track.
Figure 24 – Metallic and plastic screws
No. Description
1 Metallic screws
2 Plastic screws
NOTE:
Please, replace screws one by one to avoid spur
displacement.
Assemble the pneumatic multi connector [3] bracket to the
IM extrusion.