Safety
Electromagnetic compatibility (EMC)
The system is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the system should assure that it is used in such an environment.
Guidance and manufacturers declaration - electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1
The system uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11 Class B The system is suitable for use in all establishments,
other than domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purpose. For in-
formation purpose the system complies with
IEC61000-3-11 and is suitable for connection to pub-
lic mains network if the impedance is 0.32 Ohm or
lower
Harmonic emissions IEC
61000-3-2
Not applicable
Voltage fluctuations/ Flick-
er emissions IEC 61000-3-
3
Not applicable
Operation Manual
1000-095-051-EN Rev. 1.1 ARCOMA AB
2020-06-16 Page 33 SWEDEN