Safety
Electromagnetic compatibility (EMC)
The system is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the system should assure that it is used in such an environment.
Guidance and manufacturers declaration - electromagnetic emissions
Emissions test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Radiated emis-
sions CISPR 16-
2-3
30 MHz to 230
MHz:
QP 40
230 MHz to 1
GHz:
QP 47
30 MHz to 230
MHz:
QP 40
230 MHz to 1
GHz:
QP 47
Conducted emis-
sions CISPR 16-
2-1
150 kHz to 500
kHz:
QP 100+20,
average 90
500 kHz to 5
MHz:
QP 86+20, aver-
age 76
5 MHz to 30
MHz:
QP 90+20 (at 5
MHz) decreasing
linearly to 73+20
(at 30 MHz)
average 80 (at 5
MHz) decreasing
linearly to 60 (at
30 MHz)
150 kHz to 500
kHz:
QP 100+20,
average 90
500 kHz to 5
MHz:
QP 86+20, aver-
age 76
5 MHz to 30
MHz:
QP 90+20 (at 5
MHz) decreasing
linearly to 73+20
(at 30 MHz)
average 80 (at 5
MHz) decreasing
linearly to 60 (at
30 MHz)
Note: These limits apply to equipment with a rated power > 20 kVA and intended
to be connected to a dedicated power transformer or generator, and which is not
connected to low voltage (LV) overhead power lines. 20 dB relaxation for Quasi-
Peak (QP) is allowed for Radiography and pulsed Radiography (Intermittent
Mode).