Chapter 7 DISASSEMBLY/ASSEMBLY PROCEDURES
7 - 12
3. Using a small flat-blade screwdriver, push and hold the section of the Cover/Window
Assembly (1) near the bottom of the keypad away from the retainer (3). Immediately
apply light pressure against the keypad to initially disengage the retainer from the
Cover/Window Assembly (1).
4. Complete the removal of the PCB assemblies by inserting the screwdriver between the
top section of the Cover/Window Assembly (1) near the LCD window and the retainer
(3). Pry the Cover/Window Assembly (1) away from the retainer (3) and continue to
apply light pressure against the keypad until the PCB assemblies release and become
separated from the Cover/Window Assembly (1). Safely set the Cover/Window
Assembly aside so that neither the cover nor window become scratched.
NOTE
During reassembly of the PCB assemblies to the cover, confirm there
is no visible separation between the edges of the window and the
Cover housing. Also confirm that both the window and the LCD
screen are clean and free of lint.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals