SYSTEM CONFIGURATION Chapter 8
8 - 5
At the SELECT PROTOCOL prompt, press or to select the protocol to be
modified, and press CONFIRM to complete the selection.
Once a protocol has been selected, the pump displays NAME <protocol-name> where
<protocol-name> is the name of the selected protocol. Press CONFIRM to accept the
current protocol name, or change the name to any name that is not already assigned to
another protocol within the same library. Press CONFIRM to complete the protocol name
entry.
Next, the pump displays SELECT MODE . Press or to display the infusion
mode currently assigned to the protocol. Press <CONFIRM> to accept the current mode,
or press or as necessary to select a different mode and press CONFIRM to complete
mode selection.
NOTES
The pump offers an ALL MODES selection. If this selection is chosen,
all infusion modes will be available when the protocol is selected
during normal programming.
If ALL MODES, SINGLE DOSE, MANUAL SCHEDULE, AUTO
SCHEDULE or CUSTOM DILUTION is selected when CONFIRM is
pressed, no further settings are allowed for the protocol. If any
changes were made to the protocol, the pump displays UPDATING
immediately.
The mode assigned to a protocol will be available when that protocol is
selected during normal programming, even if the mode is not enabled
in the MODES configuration group.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
If a mode other than ALL MODES, SINGLE DOSE, MANUAL SCHEDULE, AUTO
SCHEDULE or CUSTOM DILUTION is assigned to the protocol, the pump displays
additional prompts for any of the following that apply to the mode selected:
Concentration
Dose or Rate
At each prompt, the current default value is displayed, or dashes are displayed, if no
default value is defined. Use the digits (0-9) and the decimal point (.) as necessary to
enter or modify a default value or press CLR to replace the current default value with
dashes. Press CONFIRM to complete each entry.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals