Safety Information
110 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
PATIENT TRANSFER
Warning—Use hospital safe handling protocols when transferring a patient from one surface to
another (such as the bed to a stretcher). Failure to do so could cause injury.
Progressa air surfaces—Use the Max-Inflate/P-Max Inflate surface mode to maximize the firmness of
the surface to assist in patient surface-to-surface transfers.
TRACTION EQUIPMENT
Warning—Evaluate patients for entrapment and asphyxiation risk according to facility protocol, and
monitor patients appropriately. Failure to do so could cause serious injury or death.
INTELLIDRIVE TRANSPORT SYSTEM BATTERIES
To help prevent equipment damage, obey these cautions:
• Caution—If the bed is disconnected from mains power for longer than 6 months, and the
IntelliDrive Transport System is installed and not activated
, transport system battery
performance may be affected.
– If the bed is disconnected from mains power for longer than 6 months, and the bed has
the IntelliDrive Transport System installed and not activated
, reduced battery
performance, to include the inability to charge, may result. Disconnect the bed battery
and IntelliDrive Transport System batteries for periods of storage longer than 6
months.
• Caution—If the bed is disconnected from mains power for longer than 4 days, and the
IntelliDrive Transport System is installed and activated
, transport system battery performance
may be affected.
– If the bed is disconnected from mains power for longer than 4 days, and the bed has
the IntelliDrive Transport System installed and activated
, reduced transport system
battery performance, including the inability to charge, may result.
LARGE PATIENT PRODUCT PERFORMANCE
The following bed functions may have reduced performance with patients who are near the maximum
patient weight or height for the product:
• Turn Assist—Less turn capability
• Rotation Therapy—Less turn capability
• Percussion and Vibration Therapy—Less effective
• Bed Up and Down—Slower speed while you raise the bed
• Head Up and Down—Slower speed while you raise the head section
• Knee Up and Down—Slower speed while you raise the knee section
• IntelliDrive Transport System—Slower acceleration and speed
• C-Arm Compatibility—Imaging device may not be large enough for bed and patient