Surfaces
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 79
NIGHT LIGHT
There is a night light on each side of the bed, located on the base frame. The light is on continuously
when the bed is plugged into AC power.
SURFACES
WARNING:
To help prevent injury and/or equipment damage, obey these warnings:
Warning—Some safety features of the bed may not function or may not operate as intended
with surfaces not designed specifically for this bed. Check with the surface manufacturer to
determine those safety features of the bed that have been tested and verified to work properly
with the replacement surface.
Warning—A sound risk assessment and protocol is necessary to determine the appropriate
surface for the patient's condition.
Warning—Only use Chair Egress featured Progressa Surfaces or Progressa+ Surfaces with the
Chair Egress function of the bed.
Warning—Only use StayInPlace featured Progressa Surfaces and Progressa+ Surfaces on
beds equipped with the StayInPlace Feature or reduced surface performance could occur.
Warning—The Progressa Advance Surface and Progressa Accelerate Surface weigh
approximately 60 lb (27 kg) to 70 lb (32 kg) depending which surface model you have. Always
use assistance to move the surface. Failure to do so could cause injury or equipment damage.
NOTES:
The above warning does not apply to the Progressa Prevention Surface. The Progressa
Prevention Surface has the StayInPlace feature built into the surface. The Progressa Prevention
Surface can be used on a Hillrom Progressa ICU Hospital Bed with or without the StayInPlace
feature.
Baxter recommends the use of Hill-Rom Surfaces that have been designed and tested
specifically for the bed. Customers electing to purchase replacement surfaces from other
manufacturers should confirm that the replacement surface, when used in conjunction with
the bed, meets applicable regulations, regulatory guidance and technical standards and does
not create unacceptable risks of injury to patients or caregivers. Specifically, Baxter suggests
that surfaces utilize dimensions and construction to minimize gaps where entrapment could
occur, provide for sufficient height between the surface and the top of the siderail to prevent
accidental roll-over events, provide appropriate firmness at the edges of the surface to facilitate
safe transfers into and out of the bed, and do not interfere with the proper operation of
siderails.
EQUIPOTENTIAL GROUND
The Equipotential Ground is located at the head-
end of the bed, near the power cord.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals