Caregiver Pendant Controls
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 25
CAREGIVER PENDANT CONTROLS
This section describes the pendant controls that are intended to be used only by the caregiver.
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—The caregiver pendant is for use by the caregiver only. Do not allow the patient to
use the caregiver pendant.
• Warning—The pendant is not for use inside of an oxygen tent.
The Enable control (key symbol) located on the caregiver pendant helps prevent unauthorized
operation of the controls on the pendant. The Enable control is only required for the functions
connected with the green line. The controls in the blue area do not require the Enable control to be
activated.
The Enable indicator stays on for 60 seconds. While this indicator light is on, the caregiver can activate
any pendant control.
To Activate
• Press and hold the Enable control until the indicator light
comes on. The Enable indicator light stays on for 60
seconds.
• During the 60-second period, you may activate bed
controls on the pendant without pressing the Enable
control again.
• If the enable control process is done incorrectly, the bed
will beep three times and instructions will show on the GCI
screen.
• If during the 60-second enabled time you want to turn it
off, press the Enable control. The indicator light will turn off
when the Pendant controls are no longer enabled.
To Remove from the Siderail or Footboard
• Pull straight up on the pendant.
or
• Rotate the pendant in a clockwise or counterclockwise
direction until the mount clip disengages from the
siderail or footboard.