Bed Frame Features
64 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—If the bed propels forward or reverse when depressing one of the enable switches
and not applying any pressure on either of the handles, contact your local service personnel for
repair.
• Warning—If the bed propels forward or reverse while applying pressure on either of the
transport handles and not pressing either of the enable switches, contact your local service
personnel for repair.
• Warning—If the bed is stopped on a ramp, or a patient is left unattended, set the brake to
avoid unwanted bed movement.
• Warning—Significantly reduce the speed of travel when powering the IntelliDrive Transport
System when using freestanding patient attached equipment or traveling through doorways.
Caution—The IntelliDrive Transport System is intended for indoor use only. Outdoor use may cause
temporary or permanent damage to the powered drive mechanism and/or drive belt.
WALLGUARD BUMPER SYSTEM
LINE MANAGER (P7512)
Caution—Do not wrap the power cord or communication cable around the line manager. Equipment
damage could occur.
The WallGuard Bumper System protects
the perimeter of the Hillrom Progressa
ICU Hospital Bed when it is being moved
or transported.
Roller bumpers protect the walls and
doorways when transporting the bed.
Head End Foot End
A Line Manager is on each side of the head end of the bed. The Line
Manager helps to keep lines (such as IV lines, suction lines, etc.) together
and away from the articulating frame. The flexibility of the Line Manager
lets you bend it in any direction.
Warning—Make sure the lines are not pinched or kinked and there is
sufficient slack in the lines for bed articulations and patient movement.
Failure to do so could cause injury or equipment damage.