Bed Frame Features
72 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
AUXILIARY AC RECEPTACLE OPTION
CAUTION:
Caution—Failure to store the accessory power cord when not in use, could cause damage from bed
articulation.
The receptacle option is a convenient source of AC power for accessory devices. The receptacle is not
intended for life support equipment. It is located at the foot end of the base frame.
The receptacle power cord is white, and the bed power cable is gray.
The receptacle provides up to 12 A of AC current (100 to 137 VAC beds) or 6 A of AC current
(220 to 240 VAC beds). Beds that have this option are equipped with two power cords, one for the
accessory receptacle and one for the bed. The receptacle is isolated from the bed’s AC power supply.
COMPOSER COMMUNICATION SYSTEM
The Hillrom Progressa ICU Hospital Bed is compatible with the COMposer Communication System.
With the COMposer Communication System, the bed can be monitored for the following functions:
•Bed in Low position
Siderail(s) up or down
•Brake set
Bed exit on or off
VOALTE NURSE CALL SYSTEM
The Voalte Nurse Call System is an enterprise system that connects and monitors Baxter and Hill-Rom
beds and surfaces. The system sends bed and surface data to network applications for caregivers to
view and receive alerts. For complete operational instructions for the Voalte Nurse Call System refer to
the Voalte Nurse Call System User Manual.
WARNING:
To help prevent injury and/or
equipment damage, obey these
warnings:
Warning—Do not use the
receptacle for life support
equipment. There is no
battery back-up. Plug life
support equipment directly
into facility power supply.
Warning—Do not use oxygen enriched sources near the accessory outlet.
Warning—Do not plug both power cords into the same wall receptacle. Plug the power cords
into different receptacles on separate circuits. Failure to do so can cause equipment damage or
tripping of facility power breakers.
Warning—Before you move the bed, make sure both power cords are unplugged and stored
correctly. Do not wrap the cords between the intermediate and upper frames.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Hillrom Progressa and is the answer not in the manual?

baxter Hillrom Progressa Specifications

General IconGeneral
Brandbaxter
ModelHillrom Progressa
CategoryMedical Equipment
LanguageEnglish

Related product manuals