Bed Frame Features
Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15) 65
DRAINAGE BAG HOLDERS
The holders accommodate any combination of the following drainage devices:
• Fecal incontinence bag
• 250-2000 ml Foley collection bag
• Chest drainage devices on the tan siderail holders or on the foot-end holders not in the lowest bed
height.
To help prevent injury and/or equipment damage, obey these
warnings:
• Warning—Remove drainage bags from the foot section
before you use the Chair control and remove drainage
bags from siderails before transport.
• Warning—Use caution when you position the drainage
bag tubing to keep it away from moving parts.
• Warning—Lowering the bed may cause drainage bags to
come in contact with the floor. Follow facility protocol if
they touch the floor.
• Warning—Use caution when you raise or lower a siderail
with a drainage bag present.
• Warning—Hanging a drainage bag on any part of the bed
other than the drainage hooks, without making sure
proper drainage can be achieved, may result in patient
injury.
• Warning—Remove any drainage bags from the siderails
for transport.
Caution—When you use the foot section articulations (up/down, retract/extend) or Trendelenburg or
Reverse Trendelenburg controls, make sure the drainage bags do not touch the floor.
The bed is equipped with six drainage bag holders
on each side of the bed.
Holders on the weigh frame include three (3)
holders on each side of the foot section and two
(2) hangers on each intermediate siderail.
There is one (1) green hanger on each side of the
bed that is not on the weigh frame. Only the green
drainage holders near the foot section will not be
part of patient weighing.